CertoClav EL Autoclave

CertoClav EL sterilizer - the economical solution

The economical solution

Why CertoClav EL sterilizer?

Compact: benchtop autoclave located directly at the laboratory workplace.

Fast: short cycle times thanks to direct heating

Practical: always processing the right quantity. No waiting for the big autoclaves

Reliable: often more than 20 years in use. Lifelong service

 


Technical capability

Technical capability

Power output: selectable between 125°/140°C. Special valve 115°/121°C available (0.7 - 2.7 bar / 10.2 - 39.2 psi overpressure). Solution of culture media (Agar) at 100°C

Venting: venting using flow process. Steam saturation: non-condensable gases <3% (as mandated in EN 13060)

Sterilisation material: solid instruments, liquids (culture media) bottles and devices, waste sterilisation

Size, loading: optimal usage of chamber volume thanks to loading from above. Even long instruments up to 42 cm (18 L version) can be processed

 


The 4-stage CertoSecure safety system

Safety systems

Tried-and-tested technology: 50 years of experience in autoclave construction

Multiple systems: The 4-stage CertoSecure safety system consists of an excess pressure valve, pressure valve, safety cover locking device against opening to pressure and overpressure safety recess.

Safe operation: the CertoLock lock closes safely with just one easy step. No need for time-consuming screwing as with other makes.

 


The CertoClav's construction and fittings

Type and fittings


Controller: pressure control via valve, temperature control via thermostat.

Instruments: precision manometer with drag pointer, precision thermometer

Connections: No additional water connection required. Electrics: 220V, 50-60Hz, 1900W

Standards: complies with international standards. GS mark in accordance with DIN and VDE/IEC (EN61010:1:2001 and 61010:2:040:2005). CE-conform.

Heater: integrated in base. Heating elements are not in contact with water, therefore extended service life

 


* suggested retail price. Only applicable in Germany and Austria

Safe Autoclaves

Safety first - get to know the 5-stage CertoSecure safety system… more

Get to know the CertoClav safety equipment

Practical - the autoclave fits directly on the laboratory bench

No more long walks for everyday autoclaving again... more

The table-top autoclave for your laboratory bench

No need for time-consuming screwing

The CertoClav autoclave is ready to start in 3 simple stepsmore

Ready to start in 3 easy steps

Language Selection 

Deutsch
English
Français
Español
Italiano
Polski
Türkçe