CertoSeal 200 | 300 · Rotary sealer



WelcomeThank you for choosing the CertoSeal
Since 1957, our family-owned company has been developing, manufacturing and distributing products for sterilisation and practice hygiene. Developed with in Austria.
This manual accompanies you from unpacking to annual validation. Read it completely before first use and keep it close to the device. Help: certoclav.com/sealing-devices
Contents
General information
Key facts about this manual and the manufacturer.
Please read and understand this manual before putting the sealer into operation. The device may only be used for its intended purpose. It may only be dismantled or assembled by suitably trained specialists. Safety notices, rating plates and labels must be kept complete and intact.
Symbols in this manual
Hazardous situation which, if not avoided, may result in minor to life-threatening injuries.
Situation which, if not avoided, may result in damage to the device or defective seal seams.
Important information or a tip for daily work.
Display controls are shown like this: System → Device info. Yellow numbers in figures refer to the legend next to them.
Manufacturer
CertoClav Sterilizer GmbH
Peintner Straße 10
4060 Leonding, Austria
Phone +43 732 674 278
support@certoclav.com · www.certoclav.com
Disposal
Before decommissioning, unplug the mains plug, remove all cables and take out any remaining rolls. The device contains electronics and must not be disposed of with household waste: return it to an authorised dealer, the manufacturer or a municipal collection point for waste electrical equipment.
Safety instructions
For safe operation and consistently good seal seams.
The mains cable, display and housing must not be damaged or deformed. If unusual noises, smoke, odours or uneven running occur during operation: stop immediately, unplug the mains plug and contact customer service. Do not operate in potentially explosive atmospheres.
- Unplug the mains plug before every cleaning and before every transport.
- When inserting pouches, make sure that no instruments or small parts get into the device.
- The path of the pouch must not be blocked by other objects.
- Only seal once the set temperature has been reached (status Ready).
- Keep the mains cable away from moving objects; do not kink or pinch it.
- Place the device with its rear at least 5 cm away from walls; keep the fan and ventilation slots clear.
- Do not allow any liquids to enter the device.
- Regularly remove dust from under the device and from the installation surface.
The sealing jaws reach up to 220 °C. Do not reach into the feed. Allow the device to cool down before maintenance work.
Warning sign W022 “Warning; Sharp element” according to DIN EN ISO 7010. Do not reach into the area of the autocutter while the device is switched on.
If a message appears on the display, follow the instructions. If anything is unclear or there is a risk of damage, contact the manufacturer. The device contains sensors and high-temperature elements; dismantling it without instructions is prohibited.
Device description
What the CertoSeal can do and what it is designed for.
Intended use
The CertoSeal is intended for professional use in hospitals as well as medical and dental practices. It seals surgical and medical instruments in transparent sterilisation packaging (paper/film) for subsequent sterilisation and storage. It is not a medical device within the meaning of the Medical Device Regulation. Any other use is not permitted.
Scope of delivery
- CertoSeal 200 or 300 rotary sealer
- User manual with declaration of conformity
- Mains cable
- Stainless-steel stand
Check the scope of delivery before putting the device into operation.
How it works
The CertoSeal uses fast, continuous heat-sealing technology: the pouch is drawn in automatically, sealed between ceramic heating jaws at a constant temperature and defined contact pressure, and then fed out again.
- Ready in seconds: 3 s start-up, 30 s to sealing temperature
- Temperature adjustable up to 220 °C, controlled to ± 1 °C
- Pouches are only drawn in once the set temperature is reached
- Sealing force and temperature are measured and recorded for every seal
- CertoSeal 300 with automatic cutting function: pouches cut to length, up to 5 tasks in succession


Installation and connection
Ready for operation in just a few minutes.
Unpacking
- Carefully open the transport packaging and lift the device out with two people (16 kg).
- Check the device and mains cable for transport damage. Do not put damaged devices into operation.
- Keep the packaging for later transport.
Location
- Dry, dust-protected indoor rooms only
- Outside the patient environment, at least 1.5 m from the treatment area
- Stable, level surface free from vibration
- Clearance from walls for ventilation, at least 5 cm at the rear
- No direct sunlight, no heat sources nearby
Mains connection
- 220–240 V, 50 Hz; observe the regulations of your power supplier
- Never operate with a damaged cable or plug
- Do not kink, twist, pinch or strain the cable
- Pull the plug, not the cable
- The socket must remain freely accessible so that the device can be disconnected at any time
Failure to observe the installation conditions may result in injuries, malfunctions or damage to the device.
Switching on
- Switch on the mains switch at the rear.
- The home screen appears after 3 seconds.
- Heating starts immediately; the sealing temperature is reached after about 30 seconds. Until then, the status shows Heating with the time remaining.

OperationThe home screen at a glance
Everything important at a glance: temperature, status, the data for the record and the way to all functions.
123
456
789
10111213
- Counter – consecutive number of every seal (total counter). Can only be changed with the service PIN.
- Status – Ready green, Heating · 0:27 orange with time remaining.
- Time and date
- Actual temperature – green within the set range (± 1 °C), otherwise orange.
- Set temperature – tap to change (chapter 06).
- Speed, sealing force, print length
- Batch – current batch and number of its seals.
- New – start a new batch.
- Record fields – Exp. date, AC no., Operator, Set temp., Goods, Custom. Tap a field to change it; the switch controls printing.
- Seal test – with a bell while the daily test is still due.
- Cutting (CertoSeal 300)
- Record – all seals, export to USB.
- System – settings, device info, service.
OperationSetting the sealing temperature
The correct temperature depends on the packaging material. Always follow the specifications of the manufacturer of your sterilisation packaging.

- On the home screen, tap the Set temperature.
- Choose a preset: 160, 170, 180 or 190 °C. The value is applied immediately.
- For a different value, tap Other …, enter 60–220 °C and then tap Apply.
170 °C. The tile is marked “recommended”.
Heating up
After a change, the device heats up or cools down to the new value. Status and actual temperature are shown in orange, and the time remaining is displayed. The conveyor belt only draws in pouches once the temperature has been reached.
Every change of the set temperature is automatically entered in the record (line “Set temp.” on the home screen) and cannot be edited there.

Understanding the colours
| Display | Meaning |
|---|---|
| orange | Temperature outside set value ± 1 °C – device is heating or cooling, do not seal yet |
| green | Within the set range – ready to seal |
| red | Warning after a seal outside the limits (chapter 10) |
OperationRecord fields and batch
For AC no., Operator and Goods you can store up to 12 entries each and switch between them with a single tap.


Selecting or creating an entry
- On the home screen, tap the field, e.g. Operator.
- Tick the desired entry. Exactly one entry is always active.
- To create or change an entry, tap the text, type it in and tap Apply.
Longer texts are truncated. The keyboard shows the number of characters.
Starting a new batch
A batch groups seals together, e.g. everything for one steriliser load. The Batch line shows the number and the count of seals.
- Tap New next to the batch.
- Confirm the prompt with START.
- The number increases by one and the batch’s seal count starts at 0.
The Counter at the top left, by contrast, counts every seal consecutively and is never reset.
Sealing
From the roll to a securely sealed pouch.
Loading the sterilisation roll
- Place the roll on the holding bars with the paper side facing up.
- Pull off 10 cm and guide the end under the metal plate into the insertion slot.
- Fit the limiting plates to the left and right of the roll (optional).
Use suitable pouches made of paper or paper/plastic in accordance with WS 310.1. The markings on both sides must be clearly visible.
Setting the seal seam distance
- Loosen the knurled screw under the guide plate anticlockwise.
- Slide the guide plate horizontally to the desired distance from the edge (scale 0–40 mm).
- Tighten the knurled screw again.
SealingThe sealing process step by step
If the device does not draw in the pouch, the set temperature has not yet been reached. Do not allow small parts to be drawn in: they damage the transport mechanism, the sterile goods and the packaging, and void the warranty and guarantee.
- Prepare the pouch. Choose a pouch of suitable width and length. On the CertoSeal 300, cut it with the automatic cutter (chapter 09).
- Seal the first end. Seal one end of the empty pouch.
- Insert the instruments. Pointed or irregularly shaped instruments only with tip protectors. Handles towards the peel side that will be opened. Fill level below 75 %, insert without tension.
- Check the distance. Leave at least 3 cm between the end of the instruments and the seal seam (national regulations: more than 2.5 cm), and at least 1 cm between the seal seam and the edge of the pouch.
- Insert the second end. Insert the pouch with the paper side facing down and the film side facing up as far as the guide plate. The device detects it and draws it in automatically.
- Without guide plate: insert the pouch opening flat and parallel to the front of the device and hold it throughout the entire process.
- Check the result. The seal is recorded with temperature and sealing force. If a value is outside the limits, a warning appears (chapter 10).
Only feed the packaging materials specified in this manual through the device.
Even, complete and free of creases; same colour as the factory seam at the edge.
CertoSeal 300 onlyCutting pouches automatically
Set up to five tasks with length and quantity. The CertoSeal cuts them one after the other and shows the remaining time in large digits.
123456- Current task with length and progress (2/5)
- Progress line
- Time remaining as a countdown
- Tasks 1–5: length, quantity, cut, active
- Feed, Return, Cut for individual steps
- START / STOP
How to do it
- Home screen → Cutting.
- Tap Feed to draw in the roll. Press and hold Return to feed it back.
- For each task, enter the Length (at least 80 mm) and Qty. and activate it with the tick.
- Tap START. The tasks run in sequence; after each cut, the device waits for the next one.
- STOP cancels after a confirmation prompt.
Never reach into the area of the autocutter.
Daily seal test and warnings
The CertoSeal reminds you of the test and documents the result.

At the start of each working day, a test seal seam must be produced and checked, and the result documented. As long as this has not been done today, the Seal test button on the home screen shows a bell and “Due today”.
- Tap Seal test.
- Insert a test pouch or test indicator (e.g. Seal Check) and seal it.
- Check the seal seam: visual inspection, peel test, tensile strength (chapter 15).
- Tap Passed or Failed. The result is saved in the record with the time and operator.
Do not release any pouches. Check the cause (temperature, material, contamination), repeat the test; if it fails again, contact service.
MonitoringWarning in the event of a deviation
After every seal, the CertoSeal checks temperature and sealing force. If a value is outside the limits, a warning appears immediately.

| Process value | Limit |
|---|---|
| Sealing temperature | Set value ± 5 °C |
| Sealing force | 80 N ± 16 N (64–96 N) |
| Sealing time | Heated length 50 cm ÷ belt speed; 3.0 s at 10 m/min |
What to do
- Do not release the affected pouch. Batch and counter are shown on the warning.
- Open the pouch, repack it and seal it again.
- Record shows the entry marked in red; Got it closes the warning.
- If the warning occurs repeatedly: contact service.
DIN EN ISO 11607-2 requires process variables to be monitored and deviations to be reported.
Record and export
Every seal fully documented – in the device, on USB and via RS232.
1234567- Delete – only with the service PIN and after confirmation
- Memory – up to 1000 entries; from 900 a reminder to export
- Today – number of seals, of which OK and deviations
- Status – ✓ OK, ! deviation (red border)
- Time, batch (Lot), below: counter, operator, sealing time
- Temperature and sealing force with set range, red if deviating
- Export to USB stick
Exporting to USB
- Insert a USB stick.
- Open Record and tap Export.
- Wait until the message appears (approx. 25 s), then remove the stick.
From 900 entries, a yellow notice appears. Export the record regularly.
RecordWhat is stored, sent and printed
USB export (CSV)
OK,20261104,06:02:52,0042,0000313,Bed12,180,180,80,10,3.0,175-185 64-96,763A !!,20261104,05:16:52,0042,0000312,Bed12,180,179,[101],10,3.0,175-185 64-96,079E TEST-OK,20261104,04:30:52,0042,,Bed12,,,,,,,D8BF
| Column | Content |
|---|---|
| 1 | Status: OK · !! (limit exceeded) · TEST-OK / TEST-FAIL (seal test) |
| 2–3 | Date (YYYYMMDD), time with seconds |
| 4–5 | Batch, counter |
| 6 | Operator |
| 7–9 | Set and actual temperature °C, sealing force N; values outside limits in [ ] |
| 10–11 | Speed m/min, sealing time s |
| 12–13 | Limits, checksum (proof of authenticity) |
RS232 interface
9600 baud, 8N1. After every seal, one data record for documentation systems, e.g. Segosoft. Terminated with the characters ;/r/n.
LOG:261104,20261104,20270503,00000313,Gn1,Bed12,180,Inst5,180,80;/r/n
Date, sealing date, expiry date, counter, AC no., operator, set temperature, goods, actual temperature, sealing force.
Printing on the pouch (option)
On devices with a printer: sealing date, counter and set temperature are always printed; Exp. date, AC no., Operator, Goods and Custom depending on the switches on the home screen.
Font width, text indent and 2-line print: System → Print settings.
Connecting to practice software
Transfer sealing data automatically to your practice or documentation software.
The CertoSeal 200 | 300 comes with an RS232 interface (serial port) as standard. After every seal, the device sends a data record containing the process data. Practice and documentation software, e.g. Segosoft, automatically transfers this data into the batch documentation – no copying by hand, no paper notes.
Interface
| Feature | Value |
|---|---|
| Connection | RS232 (serial) |
| Transmission | 9600 baud, 8 data bits, no parity, 1 stop bit (8N1) |
| Timing | one data record after every seal |
| Content | Date, sealing date, expiry date, counter, AC no., operator, set temperature, goods, actual temperature, sealing force |
| Format | see chapter 11, “What is stored, sent and printed” |
Connecting the CertoSeal
- Switch off the CertoSeal and the PC.
- Connect a serial cable to the RS232 port of the CertoSeal and to the PC (COM port or USB-to-serial adapter).
- Set up the interface in the practice software: 9600 baud, 8N1.
- Switch on the devices, perform a test seal and check that the data record arrives in the software.
Only connect the cable while the device is switched off. The interface only sends data; the device cannot be controlled through it.
The CertoClav IT team will help you connect to your practice software – from choosing the cable to setting up the interface. Contact: support@certoclav.com · +43 732 674 278.
On request, the CertoSeal is also available with a network interface (LAN), e.g. if the PC has no serial port or several devices are to be connected. Please get in touch.
Settings
Everything important under System.

| Setting | Meaning |
|---|---|
| Standby temperature | Temperature when not in use: room temperature, 60, 80 or 120 °C |
| Standby time | Minutes without operation after which the device switches to standby temperature: off, 10, 20, 30 min |
| Fan from | Temperature above which the fan runs |
| Language | German or English |
| Date / time | Adjust with ▲ ▼; important for a correct record |
| Brightness | Display brightness |
| Print settings | Printer, font width S/M/L, text indent, 2-line print |
| Calibration | Service only, with PIN (chapter 14) |
| Device info | Software, counter, next service, help QR code |



Service technicians onlyCalibration and service date
Calibration, service date, factory settings and deleting the record are protected by a service PIN.



Sensor correction
If the temperature measured with a reference instrument differs from the display, enter the difference as a correction. Example: set 170 °C, measured 168 °C → Temp. correction +2. The device then heats 2 °C higher. The same applies to pressure and speed; enter negative values with ±.
These values affect the validated process.
Service date
After maintenance, tap Service done (+12 months): the next date is set to today + 12 months. Once it has passed, the device shows the reminder “Service due” when switched on; Later reminds you again the next day.
Factory settings
Resets all calibration values to the delivery state, after a confirmation prompt.
Routine checks, maintenance, validation
Keeping your sealing process safe and compliant.
| When | What |
|---|---|
| Daily before work | Produce and check a test seal seam: · Visual: even, complete, without creases or defects, same colour as the factory seam. · Peel test: the paper separates from the film with hardly any residue (max. 1 cm² according to DIN EN ISO 11607-2). · Tensile strength: opening force comparable to the factory seam. · In addition, check with a suitable test device (e.g. Seal Check) and document the result on the device with Passed / Failed. |
| Regularly | Export the record to USB and archive it; remove dust from the device and installation surface. |
| Maintenance | At least every 12 months. If 30,000 sealing cycles are reached earlier, maintenance is due at that point. The device reminds you of the 12-month date; the counter is shown under Device info. |
| Annually: validation | In most German federal states (DE) and Swiss cantons (CH), initial validation and annual revalidation are mandatory for semi-critical and critical instruments, including seal strength testing according to DIN EN 868-5. |
Process variables according to DIN EN ISO 11607-2
Book an appointment or take out a maintenance and validation contract: www.certoclav.com
MaintenanceWhat to do if …
| Problem | Possible cause | Remedy |
|---|---|---|
| Pouch is not drawn in | Set temperature not yet reached | Wait until status is Ready |
| Warning “Sealing force out of range” | Packaging too thick, folded, kinked; foreign objects | Repack the pouch and seal it again; if repeated, contact service |
| Warning “Temperature out of range” | Draught, low room temperature, defective heater | Check the location, restart the device; if repeated, contact service |
| Seal seam uneven or creased | Pouch overfilled, inserted at an angle, wrong temperature | Fill level below 75 %, insert straight, temperature as specified by the manufacturer |
| Pouch burnt | Temperature too high for the material | Lower the set temperature |
| Seal test failed | Temperature, material, contamination | Eliminate the cause, repeat the test, contact service if necessary |
| “Memory almost full” | More than 900 record entries | Export the record to USB |
| “Service due” | Service date has passed | Arrange a service (www.certoclav.com) |
Care
- Before cleaning, unplug the mains plug and allow the device to cool down.
- Clean the housing with a slightly damp cloth; do not let liquid get into the device.
- Clean the display with a soft cloth; no aggressive cleaning agents.
Transport and storage
- Transport only in the original packaging or equivalent protective packaging.
- Store upright, dry and protected from dust; do not tilt.
- Ambient temperature 5–40 °C.
AppendixTechnical data
| Feature | CertoSeal 200 / 300 |
|---|---|
| Dimensions (W × D × H) | 540 × 235 × 215 mm |
| Weight | 16 kg |
| Mains voltage | 220–240 V, 50 Hz |
| Power | 500 W |
| Housing | Coated carbon steel |
| Sealing temperature | Room temperature to 220 °C, control accuracy ± 1 °C |
| Start-up time / heat-up time | 3 s to the home screen / approx. 30 s to sealing temperature |
| Seal seam width | 12 mm |
| Belt speed | ≥ 10 m/min |
| Heated length / sealing time | 50 cm / 3.0 s at 10 m/min |
| Sealing force | 80 N ± 20 %, measured |
| Edge distance | 0–40 mm |
| Cutting (300 only) | ≥ 15 m/min, width 0–330 mm, length from 80 mm |
| Operation | 7-inch colour touch display |
| Interfaces | USB (record export), RS232 (9600 baud, practice software); network (LAN) on request |
| Record memory | 1000 entries |
| Printer | Optional |
Conformity
Rotary sealer CertoSeal 200, CertoSeal 300. Applied standards: EN ISO 11607-2, EN 61010-1/-2-010, EN 61326-1, EN IEC 63000. Directives: 2006/42/EC (Machinery), 2014/35/EU (Low Voltage), 2014/30/EU (EMC), 2011/65/EU (RoHS). The EU Declaration of Conformity can be found on the next page.
AppendixEU Declaration of Conformity
The declaration is issued in German with a parallel English text.
Datum / Date: 04.11.2026
| Hersteller / Manufacturer | CertoClav Sterilizer GmbH |
| Adresse / Address | Peintner Str. 10, 4060 Leonding, Österreich |
| Produktkategorie / Product name | Durchlaufsiegelgerät / Rotary Sealer |
| Produkt Modell / Product model | CertoSeal 200, CertoSeal 300 |
EN ISO 11607-2
EN 61010-1/-2-010
EN 61326-1
EN IEC 63000
Hiermit erklären wir in alleiniger Verantwortung, dass das oben aufgeführte Produkt den Anforderungen der nachfolgenden Richtlinien sowie deren Umsetzungen in nationale Gesetze entspricht. / We hereby declare, under our sole responsibility, that the product mentioned above complies with the requirements of the following directives and their relevant transpositions into national law.
2006/42/EG (Maschinenrichtlinie) / 2006/42/EC (Machinery directive)
Konformitätsbewertungsverfahren: Richtlinie 2006/42/EG Anhang VIII / Conformity assessment procedure: Directive 2006/42/EC Annex VIII
2014/35/EU (Niederspannungsrichtlinie), 2014/30/EU (EMV-Richtlinie) und 2011/65/EU (RoHS-Richtlinie) / 2014/35/EU (Low voltage directive), 2014/30/EU (EMC directive) and 2011/65/EU (RoHS directive)
Die Konformitätsbewertung wurde unter alleiniger Verantwortung des Herstellers durchgeführt. / The conformity assessment was carried out under the sole responsibility of the manufacturer.

Michael Simon
Geschäftsführender Gesellschafter

Peintner Str. 10
A-4060 Leonding, AT
+43 732 674 278
UID Nr.: ATU22821702
certoclav.com/sealing-devices
Service and validation: www.certoclav.com
Peintner Straße 10 · 4060 Leonding · Austria
+43 732 674 278 · support@certoclav.com
English · valid from software version 2026.002 · Edition 11/2026