CertoClavCertoSeal 200 | 300 · User manual
CertoClav

CertoSeal 200 | 300 · Rotary sealer

User manual
CertoSeal 300
Ready in seconds3 s to start up · 30 s to sealing temperature
Automatic cutting functionCertoSeal 300: cut to length, up to 5 tasks
ValidatableDIN EN ISO 11607-2
MonitoredTemperature, force, sealing time
DocumentedRecord, USB, RS232
English · valid from software version 2026.002 · Edition 11/2026CertoClav Sterilizer GmbH · Leonding, Austria
CertoSeal 200 | 300 · User manualWelcome
CertoSeal 200
CertoSeal 200 · Sealing
CertoSeal 300
CertoSeal 300 · Sealing and cutting

WelcomeThank you for choosing the CertoSeal

Since 1957, our family-owned company has been developing, manufacturing and distributing products for sterilisation and practice hygiene. Developed with in Austria.

This manual accompanies you from unpacking to annual validation. Read it completely before first use and keep it close to the device. Help: certoclav.com/sealing-devices

Contents

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20262
01

General information

Key facts about this manual and the manufacturer.

Please read and understand this manual before putting the sealer into operation. The device may only be used for its intended purpose. It may only be dismantled or assembled by suitably trained specialists. Safety notices, rating plates and labels must be kept complete and intact.

Symbols in this manual

!
Warning

Hazardous situation which, if not avoided, may result in minor to life-threatening injuries.

!
Caution

Situation which, if not avoided, may result in damage to the device or defective seal seams.

i
Note

Important information or a tip for daily work.

Display controls are shown like this: System → Device info. Yellow numbers in figures refer to the legend next to them.

Manufacturer

CertoClav Sterilizer GmbH
Peintner Straße 10
4060 Leonding, Austria
Phone +43 732 674 278
support@certoclav.com · www.certoclav.com

Disposal

Before decommissioning, unplug the mains plug, remove all cables and take out any remaining rolls. The device contains electronics and must not be disposed of with household waste: return it to an authorised dealer, the manufacturer or a municipal collection point for waste electrical equipment.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20263
02

Safety instructions

For safe operation and consistently good seal seams.

!
Check before switching on

The mains cable, display and housing must not be damaged or deformed. If unusual noises, smoke, odours or uneven running occur during operation: stop immediately, unplug the mains plug and contact customer service. Do not operate in potentially explosive atmospheres.

  • Unplug the mains plug before every cleaning and before every transport.
  • When inserting pouches, make sure that no instruments or small parts get into the device.
  • The path of the pouch must not be blocked by other objects.
  • Only seal once the set temperature has been reached (status Ready).
  • Keep the mains cable away from moving objects; do not kink or pinch it.
  • Place the device with its rear at least 5 cm away from walls; keep the fan and ventilation slots clear.
  • Do not allow any liquids to enter the device.
  • Regularly remove dust from under the device and from the installation surface.
!
Hot surfaces

The sealing jaws reach up to 220 °C. Do not reach into the feed. Allow the device to cool down before maintenance work.

!
Cutting blade (CertoSeal 300 only)

Warning sign W022 “Warning; Sharp element” according to DIN EN ISO 7010. Do not reach into the area of the autocutter while the device is switched on.

i
Error messages

If a message appears on the display, follow the instructions. If anything is unclear or there is a risk of damage, contact the manufacturer. The device contains sensors and high-temperature elements; dismantling it without instructions is prohibited.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20264
03

Device description

What the CertoSeal can do and what it is designed for.

Intended use

The CertoSeal is intended for professional use in hospitals as well as medical and dental practices. It seals surgical and medical instruments in transparent sterilisation packaging (paper/film) for subsequent sterilisation and storage. It is not a medical device within the meaning of the Medical Device Regulation. Any other use is not permitted.

Scope of delivery

  • CertoSeal 200 or 300 rotary sealer
  • User manual with declaration of conformity
  • Mains cable
  • Stainless-steel stand

Check the scope of delivery before putting the device into operation.

How it works

The CertoSeal uses fast, continuous heat-sealing technology: the pouch is drawn in automatically, sealed between ceramic heating jaws at a constant temperature and defined contact pressure, and then fed out again.

  • Ready in seconds: 3 s start-up, 30 s to sealing temperature
  • Temperature adjustable up to 220 °C, controlled to ± 1 °C
  • Pouches are only drawn in once the set temperature is reached
  • Sealing force and temperature are measured and recorded for every seal
  • CertoSeal 300 with automatic cutting function: pouches cut to length, up to 5 tasks in succession
3 sfrom switching on to the home screen
30 sto sealing temperature
220 °Cmax. sealing temperature
12 mmseal seam width
CertoSeal 200
CertoSeal 300 with automatic cutting function
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20265
04

Installation and connection

Ready for operation in just a few minutes.

Unpacking

  1. Carefully open the transport packaging and lift the device out with two people (16 kg).
  2. Check the device and mains cable for transport damage. Do not put damaged devices into operation.
  3. Keep the packaging for later transport.

Location

  • Dry, dust-protected indoor rooms only
  • Outside the patient environment, at least 1.5 m from the treatment area
  • Stable, level surface free from vibration
  • Clearance from walls for ventilation, at least 5 cm at the rear
  • No direct sunlight, no heat sources nearby

Mains connection

  • 220–240 V, 50 Hz; observe the regulations of your power supplier
  • Never operate with a damaged cable or plug
  • Do not kink, twist, pinch or strain the cable
  • Pull the plug, not the cable
  • The socket must remain freely accessible so that the device can be disconnected at any time
!
Caution

Failure to observe the installation conditions may result in injuries, malfunctions or damage to the device.

Switching on

  1. Switch on the mains switch at the rear.
  2. The home screen appears after 3 seconds.
  3. Heating starts immediately; the sealing temperature is reached after about 30 seconds. Until then, the status shows Heating with the time remaining.
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20266
CertoSeal 200 | 300 · User manualOperation

OperationThe home screen at a glance

Everything important at a glance: temperature, status, the data for the record and the way to all functions.

Home screen 123 456 789 10111213
  1. Counter – consecutive number of every seal (total counter). Can only be changed with the service PIN.
  2. Status – Ready green, Heating · 0:27 orange with time remaining.
  3. Time and date
  4. Actual temperature – green within the set range (± 1 °C), otherwise orange.
  5. Set temperature – tap to change (chapter 06).
  6. Speed, sealing force, print length
  7. Batch – current batch and number of its seals.
  8. New – start a new batch.
  9. Record fields – Exp. date, AC no., Operator, Set temp., Goods, Custom. Tap a field to change it; the switch controls printing.
  10. Seal test – with a bell while the daily test is still due.
  11. Cutting (CertoSeal 300)
  12. Record – all seals, export to USB.
  13. System – settings, device info, service.
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20267
CertoSeal 200 | 300 · User manualOperation

OperationSetting the sealing temperature

The correct temperature depends on the packaging material. Always follow the specifications of the manufacturer of your sterilisation packaging.

Selecting the set temperature
  1. On the home screen, tap the Set temperature.
  2. Choose a preset: 160, 170, 180 or 190 °C. The value is applied immediately.
  3. For a different value, tap Other …, enter 60–220 °C and then tap Apply.
✓
Recommendation for CertoClav packaging

170 °C. The tile is marked “recommended”.

Heating up

After a change, the device heats up or cools down to the new value. Status and actual temperature are shown in orange, and the time remaining is displayed. The conveyor belt only draws in pouches once the temperature has been reached.

i
Set temperature in the record

Every change of the set temperature is automatically entered in the record (line “Set temp.” on the home screen) and cannot be edited there.

Understanding the colours

DisplayMeaning
orangeTemperature outside set value ± 1 °C – device is heating or cooling, do not seal yet
greenWithin the set range – ready to seal
redWarning after a seal outside the limits (chapter 10)
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20268
CertoSeal 200 | 300 · User manualOperation

OperationRecord fields and batch

For AC no., Operator and Goods you can store up to 12 entries each and switch between them with a single tap.

Entries for “Operator”; the tick marks the active one
Input with the on-screen keyboard

Selecting or creating an entry

  1. On the home screen, tap the field, e.g. Operator.
  2. Tick the desired entry. Exactly one entry is always active.
  3. To create or change an entry, tap the text, type it in and tap Apply.
!
Maximum 5 characters

Longer texts are truncated. The keyboard shows the number of characters.

Starting a new batch

A batch groups seals together, e.g. everything for one steriliser load. The Batch line shows the number and the count of seals.

  1. Tap New next to the batch.
  2. Confirm the prompt with START.
  3. The number increases by one and the batch’s seal count starts at 0.

The Counter at the top left, by contrast, counts every seal consecutively and is never reset.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/20269
08

Sealing

From the roll to a securely sealed pouch.

Loading the sterilisation roll

  1. Place the roll on the holding bars with the paper side facing up.
  2. Pull off 10 cm and guide the end under the metal plate into the insertion slot.
  3. Fit the limiting plates to the left and right of the roll (optional).

Use suitable pouches made of paper or paper/plastic in accordance with WS 310.1. The markings on both sides must be clearly visible.

Setting the seal seam distance

  1. Loosen the knurled screw under the guide plate anticlockwise.
  2. Slide the guide plate horizontally to the desired distance from the edge (scale 0–40 mm).
  3. Tighten the knurled screw again.
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202610
CertoSeal 200 | 300 · User manualSealing

SealingThe sealing process step by step

!
Only seal when “Ready”

If the device does not draw in the pouch, the set temperature has not yet been reached. Do not allow small parts to be drawn in: they damage the transport mechanism, the sterile goods and the packaging, and void the warranty and guarantee.

  1. Prepare the pouch. Choose a pouch of suitable width and length. On the CertoSeal 300, cut it with the automatic cutter (chapter 09).
  2. Seal the first end. Seal one end of the empty pouch.
  3. Insert the instruments. Pointed or irregularly shaped instruments only with tip protectors. Handles towards the peel side that will be opened. Fill level below 75 %, insert without tension.
  4. Check the distance. Leave at least 3 cm between the end of the instruments and the seal seam (national regulations: more than 2.5 cm), and at least 1 cm between the seal seam and the edge of the pouch.
  5. Insert the second end. Insert the pouch with the paper side facing down and the film side facing up as far as the guide plate. The device detects it and draws it in automatically.
  6. Without guide plate: insert the pouch opening flat and parallel to the front of the device and hold it throughout the entire process.
  7. Check the result. The seal is recorded with temperature and sealing force. If a value is outside the limits, a warning appears (chapter 10).
i
Approved materials only

Only feed the packaging materials specified in this manual through the device.

✓
A good seal seam

Even, complete and free of creases; same colour as the factory seam at the edge.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202611
CertoSeal 200 | 300 · User manualCutting

CertoSeal 300 onlyCutting pouches automatically

Set up to five tasks with length and quantity. The CertoSeal cuts them one after the other and shows the remaining time in large digits.

123456
  1. Current task with length and progress (2/5)
  2. Progress line
  3. Time remaining as a countdown
  4. Tasks 1–5: length, quantity, cut, active
  5. Feed, Return, Cut for individual steps
  6. START / STOP

How to do it

  1. Home screen → Cutting.
  2. Tap Feed to draw in the roll. Press and hold Return to feed it back.
  3. For each task, enter the Length (at least 80 mm) and Qty. and activate it with the tick.
  4. Tap START. The tasks run in sequence; after each cut, the device waits for the next one.
  5. STOP cancels after a confirmation prompt.
!
Cutting blade

Never reach into the area of the autocutter.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202612
10

Daily seal test and warnings

The CertoSeal reminds you of the test and documents the result.

Seal test with evaluation

At the start of each working day, a test seal seam must be produced and checked, and the result documented. As long as this has not been done today, the Seal test button on the home screen shows a bell and “Due today”.

  1. Tap Seal test.
  2. Insert a test pouch or test indicator (e.g. Seal Check) and seal it.
  3. Check the seal seam: visual inspection, peel test, tensile strength (chapter 15).
  4. Tap Passed or Failed. The result is saved in the record with the time and operator.
!
Failed?

Do not release any pouches. Check the cause (temperature, material, contamination), repeat the test; if it fails again, contact service.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202613
CertoSeal 200 | 300 · User manualMonitoring

MonitoringWarning in the event of a deviation

After every seal, the CertoSeal checks temperature and sealing force. If a value is outside the limits, a warning appears immediately.

Warning after a seal at 101 N
Process valueLimit
Sealing temperatureSet value ± 5 °C
Sealing force80 N ± 16 N (64–96 N)
Sealing timeHeated length 50 cm ÷ belt speed; 3.0 s at 10 m/min

What to do

  1. Do not release the affected pouch. Batch and counter are shown on the warning.
  2. Open the pouch, repack it and seal it again.
  3. Record shows the entry marked in red; Got it closes the warning.
  4. If the warning occurs repeatedly: contact service.
i
Standard

DIN EN ISO 11607-2 requires process variables to be monitored and deviations to be reported.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202614
11

Record and export

Every seal fully documented – in the device, on USB and via RS232.

1234567
  1. Delete – only with the service PIN and after confirmation
  2. Memory – up to 1000 entries; from 900 a reminder to export
  3. Today – number of seals, of which OK and deviations
  4. Status – ✓ OK, ! deviation (red border)
  5. Time, batch (Lot), below: counter, operator, sealing time
  6. Temperature and sealing force with set range, red if deviating
  7. Export to USB stick

Exporting to USB

  1. Insert a USB stick.
  2. Open Record and tap Export.
  3. Wait until the message appears (approx. 25 s), then remove the stick.
!
Memory almost full

From 900 entries, a yellow notice appears. Export the record regularly.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202615
CertoSeal 200 | 300 · User manualRecord

RecordWhat is stored, sent and printed

USB export (CSV)

OK,20261104,06:02:52,0042,0000313,Bed12,180,180,80,10,3.0,175-185 64-96,763A
!!,20261104,05:16:52,0042,0000312,Bed12,180,179,[101],10,3.0,175-185 64-96,079E
TEST-OK,20261104,04:30:52,0042,,Bed12,,,,,,,D8BF
ColumnContent
1Status: OK · !! (limit exceeded) · TEST-OK / TEST-FAIL (seal test)
2–3Date (YYYYMMDD), time with seconds
4–5Batch, counter
6Operator
7–9Set and actual temperature °C, sealing force N; values outside limits in [ ]
10–11Speed m/min, sealing time s
12–13Limits, checksum (proof of authenticity)

RS232 interface

9600 baud, 8N1. After every seal, one data record for documentation systems, e.g. Segosoft. Terminated with the characters ;/r/n.

LOG:261104,20261104,20270503,00000313,Gn1,Bed12,180,Inst5,180,80;/r/n

Date, sealing date, expiry date, counter, AC no., operator, set temperature, goods, actual temperature, sealing force.

Printing on the pouch (option)

On devices with a printer: sealing date, counter and set temperature are always printed; Exp. date, AC no., Operator, Goods and Custom depending on the switches on the home screen.

Font width, text indent and 2-line print: System → Print settings.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202616
12

Connecting to practice software

Transfer sealing data automatically to your practice or documentation software.

The CertoSeal 200 | 300 comes with an RS232 interface (serial port) as standard. After every seal, the device sends a data record containing the process data. Practice and documentation software, e.g. Segosoft, automatically transfers this data into the batch documentation – no copying by hand, no paper notes.

Interface

FeatureValue
ConnectionRS232 (serial)
Transmission9600 baud, 8 data bits, no parity, 1 stop bit (8N1)
Timingone data record after every seal
ContentDate, sealing date, expiry date, counter, AC no., operator, set temperature, goods, actual temperature, sealing force
Formatsee chapter 11, “What is stored, sent and printed”

Connecting the CertoSeal

  1. Switch off the CertoSeal and the PC.
  2. Connect a serial cable to the RS232 port of the CertoSeal and to the PC (COM port or USB-to-serial adapter).
  3. Set up the interface in the practice software: 9600 baud, 8N1.
  4. Switch on the devices, perform a test seal and check that the data record arrives in the software.
!
Note

Only connect the cable while the device is switched off. The interface only sends data; the device cannot be controlled through it.

i
Support from the CertoClav IT team

The CertoClav IT team will help you connect to your practice software – from choosing the cable to setting up the interface. Contact: support@certoclav.com · +43 732 674 278.

i
Network interface on request

On request, the CertoSeal is also available with a network interface (LAN), e.g. if the PC has no serial port or several devices are to be connected. Please get in touch.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202617
13

Settings

Everything important under System.

System
SettingMeaning
Standby temperatureTemperature when not in use: room temperature, 60, 80 or 120 °C
Standby timeMinutes without operation after which the device switches to standby temperature: off, 10, 20, 30 min
Fan fromTemperature above which the fan runs
LanguageGerman or English
Date / timeAdjust with ▲ ▼; important for a correct record
BrightnessDisplay brightness
Print settingsPrinter, font width S/M/L, text indent, 2-line print
CalibrationService only, with PIN (chapter 14)
Device infoSoftware, counter, next service, help QR code
Date and time
Print settings
Device info with help QR code
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202618
CertoSeal 200 | 300 · User manualService

Service technicians onlyCalibration and service date

Calibration, service date, factory settings and deleting the record are protected by a service PIN.

PIN entry
Calibration
“Service due” reminder

Sensor correction

If the temperature measured with a reference instrument differs from the display, enter the difference as a correction. Example: set 170 °C, measured 168 °C → Temp. correction +2. The device then heats 2 °C higher. The same applies to pressure and speed; enter negative values with ±.

!
No changes without validation

These values affect the validated process.

Service date

After maintenance, tap Service done (+12 months): the next date is set to today + 12 months. Once it has passed, the device shows the reminder “Service due” when switched on; Later reminds you again the next day.

Factory settings

Resets all calibration values to the delivery state, after a confirmation prompt.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202619
15

Routine checks, maintenance, validation

Keeping your sealing process safe and compliant.

WhenWhat
Daily before workProduce and check a test seal seam:
· Visual: even, complete, without creases or defects, same colour as the factory seam.
· Peel test: the paper separates from the film with hardly any residue (max. 1 cm² according to DIN EN ISO 11607-2).
· Tensile strength: opening force comparable to the factory seam.
· In addition, check with a suitable test device (e.g. Seal Check) and document the result on the device with Passed / Failed.
RegularlyExport the record to USB and archive it; remove dust from the device and installation surface.
MaintenanceAt least every 12 months. If 30,000 sealing cycles are reached earlier, maintenance is due at that point. The device reminds you of the 12-month date; the counter is shown under Device info.
Annually: validationIn most German federal states (DE) and Swiss cantons (CH), initial validation and annual revalidation are mandatory for semi-critical and critical instruments, including seal strength testing according to DIN EN 868-5.

Process variables according to DIN EN ISO 11607-2

Temperaturemeasured, set value ± 5 °C, in the record
Sealing forcemeasured, 80 N ± 20 %, in the record
Sealing time50 cm ÷ speed, in the record
✓
Maintenance and validation by CertoClav

Book an appointment or take out a maintenance and validation contract: www.certoclav.com

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202620
CertoSeal 200 | 300 · User manualMaintenance

MaintenanceWhat to do if …

ProblemPossible causeRemedy
Pouch is not drawn inSet temperature not yet reachedWait until status is Ready
Warning “Sealing force out of range”Packaging too thick, folded, kinked; foreign objectsRepack the pouch and seal it again; if repeated, contact service
Warning “Temperature out of range”Draught, low room temperature, defective heaterCheck the location, restart the device; if repeated, contact service
Seal seam uneven or creasedPouch overfilled, inserted at an angle, wrong temperatureFill level below 75 %, insert straight, temperature as specified by the manufacturer
Pouch burntTemperature too high for the materialLower the set temperature
Seal test failedTemperature, material, contaminationEliminate the cause, repeat the test, contact service if necessary
“Memory almost full”More than 900 record entriesExport the record to USB
“Service due”Service date has passedArrange a service (www.certoclav.com)

Care

  • Before cleaning, unplug the mains plug and allow the device to cool down.
  • Clean the housing with a slightly damp cloth; do not let liquid get into the device.
  • Clean the display with a soft cloth; no aggressive cleaning agents.

Transport and storage

  • Transport only in the original packaging or equivalent protective packaging.
  • Store upright, dry and protected from dust; do not tilt.
  • Ambient temperature 5–40 °C.
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202621
CertoSeal 200 | 300 · User manualAppendix

AppendixTechnical data

FeatureCertoSeal 200 / 300
Dimensions (W × D × H)540 × 235 × 215 mm
Weight16 kg
Mains voltage220–240 V, 50 Hz
Power500 W
HousingCoated carbon steel
Sealing temperatureRoom temperature to 220 °C, control accuracy ± 1 °C
Start-up time / heat-up time3 s to the home screen / approx. 30 s to sealing temperature
Seal seam width12 mm
Belt speed≥ 10 m/min
Heated length / sealing time50 cm / 3.0 s at 10 m/min
Sealing force80 N ± 20 %, measured
Edge distance0–40 mm
Cutting (300 only)≥ 15 m/min, width 0–330 mm, length from 80 mm
Operation7-inch colour touch display
InterfacesUSB (record export), RS232 (9600 baud, practice software); network (LAN) on request
Record memory1000 entries
PrinterOptional

Conformity

Rotary sealer CertoSeal 200, CertoSeal 300. Applied standards: EN ISO 11607-2, EN 61010-1/-2-010, EN 61326-1, EN IEC 63000. Directives: 2006/42/EC (Machinery), 2014/35/EU (Low Voltage), 2014/30/EU (EMC), 2011/65/EU (RoHS). The EU Declaration of Conformity can be found on the next page.

CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202622
CertoSeal 200 | 300 · User manualAppendix

AppendixEU Declaration of Conformity

The declaration is issued in German with a parallel English text.

CertoClav
EU-Konformitätserklärung
EU declaration of conformity
Revision / Revision: 5
Datum / Date: 04.11.2026
Hersteller / ManufacturerCertoClav Sterilizer GmbH
Adresse / AddressPeintner Str. 10, 4060 Leonding, Österreich
Produktkategorie / Product nameDurchlaufsiegelgerät / Rotary Sealer
Produkt Modell / Product modelCertoSeal 200, CertoSeal 300
Angewandte Normen / Applied standards:
EN ISO 11607-2
EN 61010-1/-2-010
EN 61326-1
EN IEC 63000

Hiermit erklären wir in alleiniger Verantwortung, dass das oben aufgeführte Produkt den Anforderungen der nachfolgenden Richtlinien sowie deren Umsetzungen in nationale Gesetze entspricht. / We hereby declare, under our sole responsibility, that the product mentioned above complies with the requirements of the following directives and their relevant transpositions into national law.

2006/42/EG (Maschinenrichtlinie) / 2006/42/EC (Machinery directive)

Konformitätsbewertungsverfahren: Richtlinie 2006/42/EG Anhang VIII / Conformity assessment procedure: Directive 2006/42/EC Annex VIII

2014/35/EU (Niederspannungsrichtlinie), 2014/30/EU (EMV-Richtlinie) und 2011/65/EU (RoHS-Richtlinie) / 2014/35/EU (Low voltage directive), 2014/30/EU (EMC directive) and 2011/65/EU (RoHS directive)

Die Konformitätsbewertung wurde unter alleiniger Verantwortung des Herstellers durchgeführt. / The conformity assessment was carried out under the sole responsibility of the manufacturer.

Leonding, 04.11.2026
Unterschrift
Michael Simon
Geschäftsführender Gesellschafter
CE
CertoClav Sterilizer GmbH
Peintner Str. 10
A-4060 Leonding, AT
support@certoclav.com
+43 732 674 278
Firmenbuchnummer: FN122912d
UID Nr.: ATU22821702
CertoClav Sterilizer GmbH · valid from software version 2026.002 · Edition 11/202623
CertoClav
Questions? We can help.
QR code help
Help and manuals
certoclav.com/sealing-devices
Service and validation: www.certoclav.com
CertoClav Sterilizer GmbH
Peintner Straße 10 · 4060 Leonding · Austria
+43 732 674 278 · support@certoclav.com
User manual CertoSeal 200 | 300
English · valid from software version 2026.002 · Edition 11/2026