CertoClav Classic
Chamber of electropolished 316L stainless steel, thermostat-controlled, nearly maintenance-free.
- Laboratory waste
- Culture media

About this model
The indestructible CertoClav Classic, the further development of the stainless steel CertoClav EL, is extremely low-maintenance. A precision thermometer, a control valve and a precision manometer with drag indicator ensure exact processes and perfect results for all applications. The model is available in two versions, each with two temperature levels.
Full description from the catalogueShow less
You can choose between the 125/140 °C and 115/121 °C models. The respective temperature level is set via the control valve in the lid. The valves cannot be interchanged as the thermostat of the appliance is matched and adjusted to the respective valve! The 2nd generation of our laboratory autoclaves comes with a new design, more functions and a newly developed touchscreen control. The steam sterilisers impress with their solid construction and are extremely low-maintenance. Venting is carried out using a flow process or fractional venting. The sterilisation chamber is made of corrosion-resistant, electropolished stainless steel (316L). This makes it extremely robust. A heater is screwed to the bottom of the chamber, but the heating elements are not directly in the water and therefore cannot calcify. This extends the service life of the autoclaves. Our laboratory autoclaves are ready to use in just a few simple steps. The sterilisation chamber can be opened and closed safely and easily with a single movement; simply turn and remove the lid. They are transportable and location-independent. Only a power supply is required. The water supply is a mixture of tap water and demineralised water. Our autoclaves are protected by a multiple safety system. It consists of a pressure relief valve, a pressure valve, a safety lid lock to prevent opening under pressure and a BI metal fuse to prevent the appliance from overheating. Important features: robust construction Temperature settings 125/140 °C Thermostat control simple operation Integrated timer venting in flow process Precision manometer with drag indicator Heating integrated in the base Ideal for the sterilisation of unwrapped solids without hollow bodies, liquids, bottles, instruments and plastics. Also suitable for culture media processing and waste sterilisation.
Intended loads
Laboratory waste
Contaminated cultures, plates and disposables, reliably inactivated.
Culture media
Agar and broth sterilised without boiling over or scorching.
Available versions
- 18 L · 125/140 °CArt. 8510174
- 18 L · 115/121 °CArt. 8510184
Prices on request. Your local dealer handles installation and the annual inspection.
What goes into the chamber.
Examples from the round wire basket that goes into this chamber.

Culture media in bottles

Plastics and rubber

Organic materials

Laboratory waste in disposal bags
Waste in bags needs a long hold. One hour is the recommendation here. A vacuum autoclave does the same load in four minutes.

Contaminated cultures and disposables
Sealed bottles
- Crimped or tightly closed glass bottles up to 40 ml: every autoclave.
- Above 40 ml, support pressure is required.
- Plastic bottles, closed or crimped, always need support pressure, whatever the volume.
- Only the VacPro 45 L offers support pressure (ordering option 8500613).
Documents
Two years guarantee on every device, on top of the statutory warranty, conditions in the terms.
Datasheet · EN
Technical datasheet
All figures on this page, the release matrix and accessories on one A4 sheet.
Catalogue page · DE
Page from the 2026 catalogue
The official catalogue entry with photography and delivery scope.
Web
Printable datasheet
The same sheet as a page, for printing straight from the browser.
Loading capacity · DE/EN
Loading capacity sheet
Chamber, carriers and how many DURAN® bottles fit per cycle.
Manuals and declarations of conformity: on request, per serial number.
What this model is released for
Straight from the CertoClav catalogue and price list. If something is not listed here, it is not listed there either. Ask us rather than assume.
Sterilisation of
- Unwrapped solids without hollow bodies
- Liquids
- Bottles
- Instruments
- Plastics
Preparation of
- Culture media and agar
Waste disposal
- Sterilisation of waste
- Waste in disposal bags
Function
- Two temperature stages
Equipment and connection
- Standalone, no plumbing needed
- Single-phase socket, no three-phase supply
CE marked. Conforms to 2014/35/EU, 2014/30/EU, 2011/65/EU.
Technical data
- Chamber volume
- 18 L
- Chamber diameter
- 26 cm
- Chamber depth
- 35 cm
- Chamber material
- Stainless steel 316L
- Max. operating pressure
- 2.7 bar
- Temperature
- 125 °C / 140 °C
- Heating power
- 1300 W
- Electrical supply
- 230 V, 50/60 Hz
- Net weight
- 8,8 kg
- Dimensions (W × D × H)
- 44.5 × 30 × 60.5 cm
- Orientation
- Vertical, top loading
- Fractionated vacuum
- No
- Flexible media probe
- No
- Drying phase
- No
- Interfaces
- :
- Country of origin
- Austria
Figures apply to article 8510174. Other versions may differ.
Loading capacity
How many laboratory bottles fit in one cycle, and which carriers the chamber takes, measured by CertoClav with DURAN® bottles.
Stainless steel series 18 L
Articles 8510184, 8510174, 8510175, 8510176
- Chamber Ø
- 260 mm
- Chamber height
- 330 mm
- Effective height
- 290 mm

Chamber

Instrument trayArt. 1300155

Wire basketArt. 1300149
Load carriers
| Item | Article | Height | Ø | Qty |
|---|---|---|---|---|
| Instrument tray | 1300155 | 41 | 250 | 1 |
| Wire basket | 1300149 | 250 | 250 | 1 |
Dimensions in mm.
Laboratory bottles per load
| Bottle | Volume | Height | Ø | Bottles |
|---|---|---|---|---|
| DURAN® Original GL 45 | 2000 ml | 265 | 136 | 1 |
| DURAN® GLS 80 | 2000 ml | 248 | 136 | 1 |
| DURAN® Original GL 45 | 1000 ml | 230 | 101 | 4 |
| DURAN® GLS 80 | 1000 ml | 218 | 101 | 4 |
| DURAN® Original GL 45 | 500 ml | 181 | 84 | 5 |
| DURAN® Youtility | 500 ml | 193 | 78 | 5 |
| DURAN® GLS 80 | 500 ml | 148 | 101 | 4 |
| DURAN® Original GL 45 | 250 ml | 143 | 70 | 9 |
| DURAN® Youtility | 250 ml | 158 | 66 | 9 |
| DURAN® GLS 80 | 250 ml | 105 | 95 | 4 |
| DURAN® Youtility | 125 ml | 124 | 55 | 12 |
| DURAN® Original GL 45 | 100 ml | 105 | 56 | 12 |
Source: CertoClav loading capacity sheet, 2022-02. DURAN® is a trademark of DWK Life Sciences.
The original sheet as PDF: Loading capacity: Stainless steel, vertical
Accessories
Parts listed for this model in the current catalogue.
Instrument plate with feet1252203The stable instrument plates with three feet and minimum water level indicator can also be used as a base plat
Service set 21310186Set gaskets for the models Classic, MultiControl 2 and Connect
Wire basket, Ø 25 cm, with handle1300149The wire basket is suitable for lifting autoclave items such as solids, waste in bags, instruments, bottles, l
Water distiller8500482High-quality, clean and demineralised water is particularly important for autoclaving. Poor water quality lead
Exhaust vapour container with hose8582020The plastic exhaust vapour container is used to collect the condensation water. The evaporation container has
Exhaust vapour hose8582016The vapour extraction hose is made of silicone and is supplied with 2 screw clamps. The hose is temperature-re
Is the CertoClav Classic the right size for your load?
Send us your typical batch: vessel type, volume, frequency. We will confirm it or point you to the model that fits better.
Ask for a dealer quote