Skip to content
Find a dealer

CertoClav EL

Valve control, a precision gauge, and sixty years of service records.

  • Laboratory waste
  • Culture media
CertoClav EL laboratory autoclave
Volume
12 L
Temperature
125 °C / 140 °C
Chamber
Aluminium 3103
Weight
6,2 kg

About this model

For more than 60 years, the CertoClav EL has been our tried and tested economic autoclave end-to-end solution with valve control. A precision thermometer, a control valve, and a precision manometer with drag indicator provide precise processes and accurate results for all applications. This model is 125/140°C , with two temperature levels.

Full description from the catalogue

These temperature settings are controlled via the control valve on the lid. These valves are not interchangeable, because the thermostat of the device is harmonized with and calibrated to the respective valve! Ideal for the sterilization of unpackaged solids without hollow bodies, liquids, bottles, instruments, synthetic materials, food preservation, culture media preparation, and waste Key features robust construction temperature levels 125/140 °C thermostat controlled simple operation high heating power electrical heater integrated in the base venting by flow procedureprecision thermometer precision manometer with drag pointer Also available as a SET with a wire basket, made of stainless steel, for convenient handling of autoclaving goods.

Intended loads

Laboratory waste

Contaminated cultures, plates and disposables, reliably inactivated.

Culture media

Agar and broth sterilised without boiling over or scorching.

Available versions

  • 12 L · 125/140 °CArt. 8501312
  • 12 L · 115/121 °CArt. 8501312A
  • 18 L · 125/140 °CArt. 8501318
  • 18 L · 115/121 °CArt. 8501318A
  • 12 L · 125/140 °C · set with wire basketArt. 8501412
  • 12 L · 115/121 °C · set with wire basketArt. 8501412A
  • 18 L · 125/140 °C · set with wire basketArt. 8501418
  • 18 L · 115/121 °C · set with wire basketArt. 8501418A

Prices on request. Your local dealer handles installation and the annual inspection.

What goes into the chamber.

Examples from the round wire basket that goes into this chamber.

  • Culture media in bottles

    Culture media in bottles

  • Plastics and rubber

    Plastics and rubber

  • Organic materials

    Organic materials

  • Laboratory waste in disposal bags

    Laboratory waste in disposal bags

    Waste in bags needs a long hold. One hour is the recommendation here. A vacuum autoclave does the same load in four minutes.

  • Contaminated cultures and disposables

    Contaminated cultures and disposables

Sealed bottles

  • Crimped or tightly closed glass bottles up to 40 ml: every autoclave.
  • Above 40 ml, support pressure is required.
  • Plastic bottles, closed or crimped, always need support pressure, whatever the volume.
  • Only the VacPro 45 L offers support pressure (ordering option 8500613).

What this model is released for

Straight from the CertoClav catalogue and price list. If something is not listed here, it is not listed there either. Ask us rather than assume.

Sterilisation of

  • Unwrapped solids without hollow bodies
  • Liquids
  • Bottles
  • Instruments
  • Plastics

Preparation of

  • Culture media and agar

Waste disposal

  • Sterilisation of waste
  • Waste in disposal bags

Function

  • Two temperature stages

Equipment and connection

  • Standalone, no plumbing needed
  • Single-phase socket, no three-phase supply

CE marked. Conforms to 2014/35/EU, 2014/30/EU, 2011/65/EU.

Technical data

Chamber volume
12 L
Chamber diameter
24 cm
Chamber depth
26 cm
Chamber material
Aluminium 3103
Max. operating pressure
2.7 bar
Temperature
125 °C / 140 °C
Heating power
1900 W
Electrical supply
230 V, 50/60 Hz
Net weight
6,2 kg
Dimensions (W × D × H)
39.5 × 27.5 × 50.5 cm
Orientation
Vertical, top loading
Fractionated vacuum
No
Flexible media probe
No
Drying phase
No
Interfaces
:
Country of origin
Austria

Figures apply to article 8501312. Other versions may differ.

Loading capacity

How many laboratory bottles fit in one cycle, and which carriers the chamber takes, measured by CertoClav with DURAN® bottles.

Aluminium series 12 L

Articles 8501312A, 8501312, 8502312A

Chamber Ø
240 mm
Chamber height
260 mm
Effective height
220 mm
  • Chamber

    Chamber

  • Instrument tray

    Instrument trayArt. 1252203

  • Wire basket

    Wire basketArt. 8583010

Load carriers

ItemArticleHeightØQty
Instrument tray1252203412251
Wire basket85830101702301

Dimensions in mm.

Laboratory bottles per load

BottleVolumeHeightØBottles
DURAN® Original GL 451000 ml2301011*
DURAN® GLS 801000 ml2181011*
DURAN® Original GL 45500 ml181845
DURAN® Youtility500 ml193785
DURAN® GLS 80500 ml1481013
DURAN® Original GL 45250 ml143707
DURAN® Youtility250 ml158668
DURAN® GLS 80250 ml105954
DURAN® Youtility125 ml1245512
DURAN® Original GL 45100 ml1055612

* 3 in the MultiControl.

Source: CertoClav loading capacity sheet, 2021-04. DURAN® is a trademark of DWK Life Sciences.

Aluminium series 18 L

Articles 8501318A, 8501318

Chamber Ø
240 mm
Chamber height
390 mm
Effective height
350 mm
  • Chamber

    Chamber

  • Instrument tray

    Instrument trayArt. 1252203

  • Wire basket

    Wire basketArt. 8583010

Load carriers

ItemArticleHeightØQty
Instrument tray1252203412251
Wire basket85830101702301

Dimensions in mm.

Laboratory bottles per load

BottleVolumeHeightØBottles
DURAN® Original GL 452000 ml2651361
DURAN® GLS 802000 ml2481361
DURAN® Original GL 451000 ml2301013
DURAN® Youtility1000 ml253934
DURAN® GLS 801000 ml2181013
DURAN® Original GL 45500 ml181845
DURAN® Youtility500 ml193785
DURAN® GLS 80500 ml1481013
DURAN® Original GL 45250 ml143707
DURAN® Youtility250 ml158668
DURAN® GLS 80250 ml105954
DURAN® Youtility125 ml1245512
DURAN® Original GL 45100 ml1055612

Source: CertoClav loading capacity sheet, 2021-04. DURAN® is a trademark of DWK Life Sciences.

The original sheet as PDF: Loading capacity: Aluminium, vertical

Get to know CertoClav EL autoclavesen

See it run

The video opens at YouTube only when you press play. Nothing is loaded from there before that.

Is the CertoClav EL the right size for your load?

Send us your typical batch: vessel type, volume, frequency. We will confirm it or point you to the model that fits better.

Ask for a dealer quote
DatasheetFind a dealer